For unparalleled quality, trust the results with Whirl-Pak for your sampling needs. Learn more here.

METROHM 61006510

Metrohm USA
Metrosep A Supp 5 - 100/4.0


Column Type:
 
IC Anion Column
Length:
 
100mm
For Use With:
 
_
Carrier Material:
 
Polyvinyl Alcohol with Quarternary Ammonium Groups
Eluents:
 
Carbonate eluent (standard eluent), Cola eluent
Temperature Range:
 
20 to 60ºC
Metrohm™ Metrosep A Supp 5 Separation Columns

The Metrosep A Supp 5 Anion Columns allows highly efficient, rapid separations. This property makes the Metrosep A Supp 5 the standard column for short analysis times and for determination of anions which elute late (e.g., perchlorate).

Carbonate eluent (standard eluent)

  • Sodium hydrogen carbonate: 1.0 mmol/L; 168 mg/2 L
  • Sodium carbonate: 3.2 mmol/L; 678 mg/2 L

Chromate eluent

  • Sodium hydrogen carbonate: 4.0 mmol/L; 672 mg/2 L
  • Sodium carbonate: 12.8 mmol/L; 2712 mg/2 L

Cola eluent

  • Sodium hydrogen carbonate: 3.0 mmol/L; 504 mg/2 L
  • Sodium carbonate: 7.0 mmol/L; 1484 mg/2 L

Metrohm USA
Column Type:
 
IC Anion Column
Length:
 
100mm
For Use With:
 
_
Carrier Material:
 
Polyvinyl Alcohol with Quarternary Ammonium Groups
Eluents:
 
Carbonate eluent (standard eluent), Cola eluent
Temperature Range:
 
20 to 60ºC
Standard Flow:
 
0.7 mL/min
pH Range:
 
3 to 12
Maximum Flow Rate:
 
0.8mL/min
Housing Material:
 
PEEK
ID:
 
4mm
Column Capacity:
 
32 µmol (Cl–)
Maximum Pressure:
 
15MPa
Particle Size:
 
5 μm