SIGMA 20511
General Description
The liner is the secound most important part of the inlet. It is the centerpiece of the inlet syste, basically a glass tube between the septum and the column, where the sample is evaporated and brought into the gas phase. There are three most important parameters one should keep in mind while choosing a liner, firstly liner volume, considering a typical liner has a volume of 900µl, secondly liner treatment, one should always use a deactivated liner, so that the glass does not react with the analytes introduced in the liner, thirdly liner design, different designs are available like straight, tapered, baffled, and without glass wool and one should be careful enough while choosing the design such that it is compatible with the desired injection mode.
Deactivated (silanized) glass inlet liner for packed injection ports.
- Fits Agilent (5890, 6890, and 7890)
Application
Inlet liner maybe used for the determination and evalution of volatile components in food products. the vegetable oil maybe directly placed into the GC inlet liner packed with glass wool and thhen inserted into the heated inlet of the GC for analysis.