HACH LPV442.99.01022
The platform employs a patented optical design that sees more of your sample than any other turbidimeter, delivering the best low level precision and sensitivity while minimizing variability from test to test. For the first time you will be able to remove the uncertainty of which measurement to trust, thanks to identical 360° x 90° Detection Technology in both instruments. The TU5 Series dramatically reduces the time needed to get a turbidity measurement you can rely on. The TU5200 Lab Turbidimeter, with the 360° x 90° Detection Technology, sees more of your sample meaning you no longer have to apply silicone oil to your vial and index to ensure an accurate reading. In addition, you can have confidence that you will get consistent results in the lab, from operator to operator, since the impact of small interferences is minimized due to the collection of light at 360°.
Only the new TU5 Series Lab & Process Turbidimeters with 360° x 90° Detection deliver unprecedented confidence that a change in your reading is a change in your water.
Specifications:
Accuracy: ±2 % plus 0.01 NTU from 0 - 40 NTU; ±10 % of reading from 40 - 1000 NTU based on Formazin primary standard (at 25 °C)
Calibration Method:
For Formazin and Stablcal:
20 NTU from 0 to 40 NTU; at 20 FNU and 600 NTU for full range
Custom calibration for up to 6-point calibrations
Degrees: 0 to 100mg/L
SDVB 0 - 1000 NTU
Certifications:
CE compliant
US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version
Complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Australian ACMA Marking
Communication: Link2sc
Dimensions (H x W x D): 195mm x 409mm x 278mm
Display: 7in WVGA (800 x 480) color TFT-LCD with touch screen
Enclosure Construction: PC/ABS Blend Plastic
Light Source: Class 2 laser product, with embedded 650 nm (EPA 0.43 mW) or Class 1 laser product, with embedded 850 nm (ISO), max. 0.55 mW (complies with IEC/EN 60825-1 and to 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Model: TU5200 ISO
Operating Humidity: 80 % at 30°C (non condensing)
Operating Temperature Range: 10 to 40°C
Range:
EPA:
0 - 700 NTU / FNU / TE/F / FTU
0 - 100 mg/L
0 - 175 EBC
ISO:
0 - 1000 NTU / FNU / TE/F / FTU
0 - 100 mg/L
0 - 250 EBC
Regulatory: DIN EN ISO 7027
Repeatability: <40 NTU: Better than 1% of reading or ±0.002 NTU on Formazin at 25 °C (77 °F), whichever is greater
>40 NTU: Better than 3.5% of reading on Formazin at 25 °C (77 °F)
Resolution: 0.0001 NTU / FNU / TE/F / FTU / EBC / mg/L
Response Time:
Signal averaging off: 7 seconds
Signal averaging on: 10 seconds (when averaging time is 5 seconds)
Sample Temperature: 4 to 70°C
Source Lamp: 850 nm Class 2 Laser
Storage Conditions: -30 to 60°C (-22 to 140°F)
Stray Light: <10 mNTU
Units:
NTU, FNU, TE/F, FTU, EBC;
mg/L if calibrated with Degrees calibration curve
Weight 2.4kg (5.29lb)